Evaluating the Knowledge of Endotracheal Cuff Pressure Monitoring Among Critical Care Providers by Palpation of Pilot Balloon and By Endotracheal Tube Cuff Manometer



Abstract

Introduction

Mishandled endotracheal cuff pressure may either make ventilation difficult or cause damage to the airway. Therefore, the aim of this audit was to assess the knowledge about endotracheal cuff pressure monitoring with a manometer and manual palpation of pilot balloon among critical care providers.

Methods

This audit includes 150 critical care providers having experience of handling endotracheal tube (ETT) cuff at critical care area of National Institute of Cardiovascular Diseases (NICVD), Karachi from April 2017 to June 2017. Knowledge about endotracheal cuff pressure monitoring with the manometer and deleterious effects of mishandled ETT cuff was assessed using a self-reported questionnaire. Enrolled healthcare providers were asked to palpate the patient and cuff pressure was recorded and categorized.

Results

Out of 150 participants, 66 (44.0%) were doctors. Only 46 (30.67%) participants had prior knowledge about ETT cuff manometer and 110 (73.33%) had never used a manometer. Similarly only 42 (28.0%) had knowledge of hazardous effects of mishandled ETT cuff. Kappa coefficient of 0.155 with p=0.015 showed significant yet low agreement between participant prediction and the actual amount of air in cuff balloon. Agreement level was comparatively higher for staff as compared to doctors with a Kappa coefficient of 0.210 (p=0.018) vs. 0.133 (p=0.099).

Conclusion

In this study of knowledge and practice of ETT tube cuff pressure monitoring, we observed low levels of knowledge (30.67%), poor adherence to standard practice (73.33%) and were able to demonstrate poor agreement (Kappa coefficient 0.155; p=0.015) between the palpation method and cuff manometer measurements for assessing cuff pressure.

Introduction

Endotracheal tube cuff pressure monitoring is an integral component of ICU (Intensive Care Unit) care. High cuff pressures have been reported to directly cause airway complications; epithelial necrosis (with fistula formation), tracheomalacia, laryngeal inflammation, and stenosis. Low cuff pressures may contribute to aspiration and are a direct cause of ineffective ventilation [1-4]. The 2011 guidelines from the American Society of Anesthesiologists recommend endotracheal cuff pressure monitoring by a manometer immediately after intubation and throughout the course of mechanical ventilation [5]. High volume, low pressure cuffed endotracheal tubes are currently in use and the intra-cuff pressure is almost equal to the pressure it exerts on tracheal mucosa. There is no consensus on what is an appropriate cuff pressure, cuff pressure range between 25 to 40 cm H2O has been suggested [6], but we know that tissue perfusion pressure is more than or equal to 25cm H2O and higher pressure will compromise tissue integrity. Tracheal ischemia, ulceration, inflammation, stenosis at the site of the injured tracheal wall, and granulation may develop as a result of excessive cuff pressures [7]. In practice, these complications can be avoided by maintaining optimal cuff pressure.

A cuff pressure greater than 34 cm H2O can result in decreased perfusion of the tracheal wall, whereas total obstruction of blood flow to the tracheal wall can occur at a compression pressure of 50 cm H2O. In fact, cuff pressure of 27 cm H2O may reduce blood flow to the cuff site by 75%. A minimal cuff pressure of 20 cm H2O has been recommended for positive-pressure ventilation and prevention of aspiration [7]. Different studies have shown the complication related to under-inflation and over-inflation of endotracheal cuff pressures. A prospective study assessing the risk factors for tracheal stenosis in patients subjected to prolonged tracheal intubation (more than eight hours in duration) concluded that strict monitoring of the cuff pressure thrice a day might help to prevent ischemic lesions and tracheal stenosis [8].

Routine monitoring of cuff pressures by manometry is now standard practice in most intensive care unit (ICU) and operating rooms (OR). Unfortunately, in Pakistan, this is infrequent. Part of the problem is a lack of awareness about the complications of cuff over or under inflation. Other factors are deficient knowledge regarding the limits of cuff pressure and monitoring methods [9]. In most occasions, endotracheal tube cuff pressures are ‘guesstimated’ by palpating the pilot balloon instead of directly measured. The objective of this study was to audit knowledge of critical care providers about manometric endotracheal cuff pressure monitoring and to compare manometric monitoring with the palpation method.

Materials & Methods

With the approval of the Institutional Ethical Review Committee, this audit was conducted in critical care areas, coronary care unit (CCU), intensive care unit (ICU), and emergency room (ER), of National Institute of Cardiovascular Diseases (NICVD), Karachi from April 2017 to June 2017. Critical care providers with at least six months experience of handling endotracheal tube (ETT) cuffs were enrolled for the study. Verbal consent for participation was obtained. On a predefined structured questionnaire, data regarding participant demographics, education, and professional experience was collected. Knowledge about manometric endotracheal cuff pressure monitoring and the deleterious effects of mishandled ETT cuff were assessed using a self-reported questionnaire. In intubated patients with already inserted ET tubes (internal diameters of 7.0 mm, 7.5 mm, and 8.0 mm), cuff pressures were measured using a Portex manometer (Smith Medical International, Minnesota, US) and then ‘estimated’ by the palpation of the pilot balloon. The cuff pressure was recorded and categorized as under-inflated (less than 20 cm H2O), over-inflated (higher than 30 cm H2O) or within the range (20 to 30 cm H2O). If the ETT cuff pressures were out of range, it was optimized by cuff manometer. To minimize bias, healthcare professionals not directly involved in that patient’s care were asked to assess cuff pressures.

IBM SPSS Statistics for Windows, Version 21.0 (IBM Corp., Armonk, NY, US) and SAS (version 8.1, SAS Institute, Inc., Cary, US) was used to analyze the data. Frequency and percentages were calculated for categorical variables. Chi-square test was applied to test for associations between categorical variables. Kappa coefficient was calculated to assess the agreement between the predicted and actual amount of air in the cuff balloon. Two-sided p-value ≤ 0.05 was taken as criteria for statistical significance.

Results

Out of 150 participants, 66 (44%) were doctors, 71 (47.33%) of the participants had more than five years of experience. Only 46 (30.67%) of the participants had prior knowledge about ETT cuff manometry and 110 (73.33%) had never used a manometer. Similarly, only 42 (28%) of the participants had knowledge of the hazardous effects of mishandled ETT cuff pressures. Baseline characteristics of the study participants are summarized in Table 1.

Characteristics Frequency (%)
Designation
Doctors 66 (44%)
Staff 84 (56%)
Shift of health care personnel
Morning 103 (68.67%)
Evening 40 (26.67%)
Night 7 (4.67%)
Year of experience
1-2 years 22 (14.67%)
>2-5 years 57 (38%)
> 5 years 71 (47.33%)
Knowledge about ETT cuff manometer
Yes 46 (30.67%)
No 104 (69.33%)
Use of manometer
More often 24 (16%)
Less often 10 (6.67%)
Rarely 6 (4%)
Never 110 (73.33%)
Knowledge of hazardous effects of mishandled ETT cuff
Yes 42 (28%)
No 108 (72%)
Participants knowledge of air required to fill cuff balloon in mmHg
0-2 3 (2%)
2-4 26 (17.33%)
5-8 47 ´(31.33%)
8-10 28 (18.67%)
10-15 10 (6.67%)
>15 1 (0.67%)
Not known 35 (23.33%)

Almost half of the participating critical care providers, 68 (45.33%), never did routine palpation of the ET cuff. Only 67 (44.67%) participants correctly estimated the amount of air in the cuff balloon as within the range, whilst, manometric assessment of air was within range in 58 (38.67%) cases. Knowledge assessment of the amount of air in cuff balloon of the study participants is summarized in Figure 1.

Overall Kappa coefficient was found to be 0.155 with p-value <0.05, representing a statistically significant but low agreement between the actual and estimated amount of air in cuff balloon. Nurses’ prediction of air in ET cuff balloon showed relatively more agreement than that of doctors’, 0.210 vs. 0.133. Kappa coefficient and 95% confidence integral for agreement between predicted and actual air in the cuff balloon is presented in Table 2.

  Total (n=150) Designation
Doctors (n=66) Staff (n=84)
Kappa Coefficient 0.155 0.133 0.210
95% Lower Conf Limit 0.022 -0.042 0.043
95% Upper Conf Limit 0.288 0.309 0.389
p-value 0.015* 0.099 0.018*

Discussion

Our study shows a surprisingly low level of knowledge amongst critical care providers regarding endotracheal tube (ETT) cuff pressure measurement by manometry; 46 (30.67%). A majority, 73.33% (110), had never used a manometer and very few, 42 (28%), were aware of the potentially harmful effects of inappropriate (either high or low) ETT cuff pressure. Despite various innovations and developments in recent years maintaining ETT cuff pressure at its optimal level remains one of the major challenges in the critical care setting [10]. ETT cuff pressure between 20 to 30 cm H2O is the recommended safe range [11], over-filling or under-filling of ETT cuff may result in various injuries and complications [12-14]. Over-inflation of ETT cuff is reported to be associated with alterations in swallowing, tracheoesophageal fistulae, stenosis and necrosis, tracheal wall ischemia, tracheal rupture, and recurrent laryngeal nerve palsy. It is also reported to be more commonly associated with a sore throat and stridor in patients after extubation [11-13]. Similarly, ventilator-associated pneumonia is associated with under-inflation of ETT cuff due to aspiration of pharyngeal secretions [12,14].

Owing to the fact that mishandling of ETT cuff pressure and under or over inflation may result in various complications in patients after intubation, which may result in increased hospitalization and healthcare burden in these patients. A number of pressure monitoring techniques employed in clinical practice are described in the literature which includes cuff pressure measurement, minimal occlusive volume, and minimal leak technique [13,14]. Cuff pressure measurement using aneroid manometer was found to be associated with lesser complications in patients after intubation [15]. Unfortunately, a significant number of critical healthcare professionals, 45.33% (68), in our sample never did the routine palpation of cuff and majority of them 73.33% (110), have never used a manometer.

Educating critical health care providers regarding the importance of monitoring cuff pressures and use of manometer is mandatory if we are to improve the delivery of our healthcare services and to reduce iatrogenic morbidity and mortality in critical illness.

Conclusions

In this study of knowledge and practice of ETT tube cuff pressure monitoring, we observed low levels of knowledge, poor adherence to standard practice and were able to demonstrate poor agreement between the palpation method and manometric method for assessing cuff pressure. We recommend every institution should evaluate their own strategy and develop protocols to ensure optimal delivery of safe healthcare in the ICU.


References

  1. Klainer A, Turndorf H, Wu W, Maewal H, Allender P: Surface alterations due to endotracheal intubation. Am J Med. 1975, 58:674-683. 10.1016/0002-9343(75)90504-5
  2. McHardy F, Chung F: Postoperative sore throat: cause, prevention and treatment. Anaesth. 1999, 54:444-453. 10.1046/j.1365-2044.1999.00780.x
  3. Sadaka F, Trottier S, Kane T: The cuff-leak test and post-extubation stridor in intensive care unit patients. Crit Care Med. 2004, 32:111-119.
  4. Stauffer JL, Olson DE: Tracheal stenosis complicating soft-cuff endotracheal tube intubation and tracheotomy: results of a prospective study. Crit Care Med. 1978, 6:96.
  5. Sultan P, Carvalho B, Rose BO, Cregg R: Endotracheal tube cuff pressure monitoring: a review of the evidence. J Perioper Pract. 2011, 21:379-386. 10.1177/175045891102101103
  6. Vyas D, Inweregbu K, Pittard A: Measurement of tracheal tube cuff pressure in critical care. Anaesth. 2002, 57:275-277. 10.1046/j.1365-2044.2002.2404_3.x
  7. Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J, Wadhwa A: Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol. 2004, 4:8. 10.1186/1471-2253-4-8
  8. Maboudi A, Abtahi H, Hosseini M, Tamadon A, Safavi E: Accuracy of endotracheal tube cuff pressure adjustment by fingertip palpation after training of intensive care unit nurses. Iran Red Crescent Med J. 2013, 15:381-384. 10.5812/ircmj.4164
  9. Pia Hedberg R, Carolina Eklund R, Sandra Högqvist R: Identification of a very high cuff pressure by manual palpation of the external cuff balloon on an endotracheal tube. AANA J. 2015, 83:179-182.
  10. Lizy C, Swinnen W, Labeau S, Blot S: Deviations in endotracheal cuff pressure during intensive care. Am J Crit Care. 2011, 20:421-422. 10.4037/ajcc2011398
  11. Labeau SO, Bleiman M, Rello J, Vandijck DM, Claes B, Blot SI: Knowledge and management of endotracheal tube cuffs. Int J Nurs Stud. 2015, 52:498-499.
  12. Lorente L, Blot S, Rello J: New issues and controversies in the prevention of ventilator-associated pneumonia. Am J Respir Crit Care Med. 2010, 182:870-876. 10.1164/rccm.201001-0081CI
  13. Jordan P, Van Rooyen D, Venter D: Endotracheal tube cuff pressure management in adult critical care units. South Afr J Crit Care. 2012, 28:13-16.
  14. Rose L, Redl L: Survey of cuff management practices in intensive care units in Australia and New Zealand. Am J Crit Care. 2008, 17:428-435.
  15. Liu J, Zhang X, Gong W, et al.: Correlations between controlled endotracheal tube cuff pressure and postprocedural complications: a multicenter study. Anesth Analg. 2010, 111:1133-1137. 10.1213/ANE.0b013e3181f2ecc7
Original article
peer-reviewed

Evaluating the Knowledge of Endotracheal Cuff Pressure Monitoring Among Critical Care Providers by Palpation of Pilot Balloon and By Endotracheal Tube Cuff Manometer


Author Information

Jawed Abubaker

Internal Medicine, Dr. Ziauddin University and Hospital, Karachi, PAK

Syed Zia Ullah

Adult Cardiology, National Institute of Cardiovascular Diseases, Karachi, PAK

Shazia Ahmed

Adult Cardiology, National Institute of Cardiovascular Diseases, Karachi, PAK

Aziz U. Rehman Memon

Critical Care Unit, National Institute of Cardiovascular Diseases, Karachi, PAK

Zohaib J. Abubaker

Medical Education, Dr. Ziauddin University and Hospital, Karachi, PAK

Muhammad Imran Ansari

Critical Care Unit, National Institute of Cardiovascular Diseases, Karachi, PAK

Musa Karim Corresponding Author

Statistics, National Institute of Cardiovascular Diseases, Karachi, PAK


Ethics Statement and Conflict of Interest Disclosures

Human subjects: Consent was obtained by all participants in this study. National Institute of Cardiovascular Diseases Karachi issued approval ERC-12/2017. Approval was obtained for the ethical review committee of the National Institute of Cardiovascular Diseases (NICVD), Karachi . Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue. Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following: Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work. Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work. Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.