Technical report
peer-reviewed

Turtleheading the Tough Aortic Necks! A Novel Endovascular Method to Avoid Bird-beaking, Invagination, and Stent Migration in Thoracic Aortic Grafts



Abstract

Conventional repair of aortic pathology such as thoracic aortic aneurysms (TAA), aortic dissections, and intramural hematomas (IMH) involves major cardiothoracic surgery. Complication rates can be as high as 30%, therefore percutaneous endograft placement has become the new gold standard. However, not every patient is a suitable candidate for endovascular repair of a thoracic aneurysm, especially, patients with a very short proximal landing zone neck or a difficult type II or type III configuration of the aortic arch. Emerging techniques have been described in the literature, but until now none have been able to confidently conquer this problem. Stacked stents in a “turtlehead” fashion offer a solution to this obstacle. The turtlehead technique utilizes commercially available stents deployed in an on-label fashion to create a rigid yet conformable endograft that can precisely treat difficult proximal landing zone necks.

Introduction

Approximately 10% of thoracic aortic aneurysms (TAA) occur in the aortic arch and 40% occur in the descending aorta [1]. The majority of TAA are asymptomatic given the nature of their indolent course. The five-year survival rate of untreated TAA measuring at least 6 cm is approximately 54%, therefore treatment is imperative. The size of the aneurysm, the arch anatomy, and proximal landing zone length are important factors to be considered during endovascular aortic stent grafting for the repair of TAA. There are three different types of aortic arch which are delineated based on the vertical distance from the innominate artery and top of the aortic arch [2]. Type 1 arch is characterized by a distance of <1 diameter of the left common carotid artery (CCA). Type 2 vertical distance measures between 1-2 diameters of CCA, and type 3 represents diameters >2 (Figure 1).

 

Stent-grafts that do not conform to the contours of the aortic arch can extend into the lumen at the inner curvature of the arch, forming a bird-beak on imaging [3]. Patients with type 3 arch anatomy present a complex repair profile due to the concern of bird-beaking, invagination, and stent migration. Bird-beaking is a term that is used to describe when a stent does not have good wall apposition proximally allowing pressure and force on the outside of a stent instead of on the inner lumen. Bird-beaking can lead to invagination in which the proximal portion of the stent infolds on itself leading to abrupt narrowing or complete occlusion. Approximately 1.5 cm landing zone is needed for adequate positioning and covering of the aneurysm with an endovascular graft. If the aneurysm is in close proximity to the origin of the left subclavian artery, for adequate seal of the landing zone, coverage of the left subclavian artery must be performed. Coverage of the origin of the left subclavian artery for adequate proximal landing zone occurs in approximately 35% of cases [4-6]. This article is presented to illustrate a novel technique named “turtleheading” for the treatment of complex aortic pathology. The turtlehead technique is utilized for rigidity, conformability, and precision. Although the authors have successfully performed this technique on multiple patients, three demonstrative examples (aneurysm, dissection, and traumatic injury) of patients with difficult arch anatomy and short landing zones treated with endovascular graft repair using the turtlehead technique are shown.

Case 1 is an 89-year-old male with history of congestive heart failure, coronary heart disease, pulmonary emphysema, hypertension, and atrial fibrillation. A computed tomography angiography (CTA) of the chest was obtained which showed a massive thoracic aortic aneurysm measuring up to 9 cm in diameter and a type III arch with a short proximal landing zone neck. Endovascular aortic stent grafting with a single graft was not feasible in our case due to concern of migration and the possible complication of bird-beaking, which can lead to invagination and a disastrous outcome.

Case 2 is a 75-year-old female with history of hypertension, chronic obstructive pulmonary disease, lung cancer, and chronic pulmonary emboli presented with acute onset back pain. A CTA of the chest demonstrated a large intramural hematoma (dissection) from the level of the aberrant right subclavian artery to the celiac trunk.

Case 3 is a 61-year-old male with history of coronary artery disease and left-sided pacemaker. He was riding a bicycle and was struck by a vehicle. A trauma thoracic CTA was performed showing an aortic tear with pseudoaneurysm formation just distal to the left subclavian artery with only a 5.9 millimeter (mm) landing zone proximal neck along the lesser curvature of the aortic arch.

Technical Report

For the aneurysm case presented, percutaneous left brachial arterial access was performed using the micropuncture system and upgraded to a 6 French vascular sheath. Percutaneous right common femoral arterial access was performed using the micropuncture system after a groin cut down by the vascular surgical team and upgraded to a 24 French Dry Seal sheath (W L Gore and Associates, Flagstaff, AZ, USA). An angiogram was performed with a flush catheter from a left brachial approach showing a massive thoracic aortic aneurysm measuring 9 cm and a type III arch with a short proximal landing zone neck (Figure 2).

 

Due to the descending thoracic aortic aneurysm size, location, and subsequent short landing zone, there is increased likelihood of bird-beaking and endoleak. Therefore, the decision was made to deploy multiple thoracic aortic stent grafts in a turtlehead overlapping fashion beginning distally to proximally. The distal portion of the aortic treatment zone measured 28 mm in diameter and the proximal portion of the native aorta measured 38 mm. Therefore, through the right common femoral sheath, five thoracic aortic stent/grafts [34 mm x 34 mm x 10 cm, 37 mm x 37 mm x 10 cm, 40 mm x 40 mm x 10 cm, 45 mm x 45 mm x 10 cm, and 45 mm x 45 mm x 10 cm] (W L Gore and Associates, Flagstaff, AZ, USA) were sequentially deployed in the aforementioned turtlehead overlapping fashion. The amount of overlap increases with each subsequent stent from distal to proximal. The final stent is placed with 85-90% overlap with only approximately a centimeter of stent deployed proximally for exact precision known as the turtlehead (Figure 3).

 

This technique provides rigidity to the stent grafts which improves accuracy of placement and decreased risk of movement of the final stent/graft placement, bird-beaking or endoleak. The final endograft was positioned proximal to the origin of the left subclavian artery and uneventfully deployed. Subsequently, balloon angioplasty with a trilobed balloon (W L Gore and Associates, Flagstaff, AZ, USA) was performed at the proximal and distal ends of the grafts. Aortography was then performed through pigtail flush catheter placed in the thoracic aortic arch which demonstrated complete coverage of the descending thoracic aortic aneurysm (Figure 4).

 

No endoleak or bird-beaking was observed. At this time, angiography of the left subclavian and vertebral arteries was performed. A 14 mm Amplatzer vascular plug (Abbott, St. Paul, MN, USA) was deployed at the origin of the left subclavian artery with preservation of the left vertebral artery. The left brachial arterial access site was closed using manual pressure. The vascular surgical team then surgically closed the right common femoral artery access site and groin region. A glass model was used for additional explanation of the technique (Figures 5-10).

 
 
 
 
 
 

For the intramural hematomas (IMH) case presented, the aorta diameter measures 37 mm just distal to the aberrant right subclavian artery and 32 mm at the level of the celiac artery. The initial graft placed was 40 mm x 15 cm and landed just above the celiac artery in order to begin the foundation for subsequent support. The second graft, which was placed for rigidity and conformity around the aortic arch was 40 mm x 15 cm. This graft was deliberately placed just short of the target to allow realestate for the final turtlehead device. A significant bird-beak was observed. Because of the nature of the disease process being an IMH, angioplasty was contraindicated. The final graft (40 mm x 10 cm) known as the turtlehead was precisely placed uneventfully just distal to the aberrant right subclavian artery orifice. This was placed with approximately one centimeter of graft material exposed. This provided circumferential wall apposition which completely eliminated the previously seen bird-beak (Figures 11-20).

 
 
 
 
 
 
 
 
 
 

For the trauma case presented, the aorta diameter measures 24 mm just distal to the left subclavian artery (LSA). There is a very short proximal landing zone neck from LSA to the arotic injury. The initial graft placed was 26 mm x 10 cm (W L Gore and Associates, Flagstaff, AZ, USA) and landed approximately 10 mm from the LSA in order to begin the foundation for subsequent support. This graft was deliberately placed just short of the target to allow real-estate for the final turtlehead device. The second and final turtlehead graft, which was placed for precision, was a 28 mm x 10 cm (W L Gore and Associates, Flagstaff, AZ, USA). Angioplasty was contraindicated in this traumatic injury case. Final angiogram showed excellent results with exclusion of the aortic tear and no bird beak (Figures 21-23).

 
 
 

Discussion

The management and treatment of aortic pathology has undergone significant changes in recent years with the advent of less invasive techniques. Preliminary data has shown less postprocedural complications and mortality with endograft repair in comparison to open repair [7]. Nationwide, more institutions are therefore favoring endovascular repair as first-line option for TAA treatment. Approximately 14% of patients with TAAs greater than 6 cm in diameter have annual risk of rupture or dissection [8]. Therefore, when the aneurysm size is greater than 6.5 cm in an average risk patient, operative repair is recommended [4]. Repair is also recommended if the aneurysm is growing more than 1 cm per year. TAA can be treated with open surgical repair or with endovascular stent grafting. The 30-day mortality of endovascular repair of an unruptured TAA is 5.2% compared to 12% with open surgical repair [9]. Overall, including complication rates and post-treatment re-intervention rates, the endovascular stent grafting of TAA is cost-effective in comparison to the open repair [7]. Therefore, endovascular repair is the favored treatment option in our patients. Traditionally with endovascular approach, a single body graft is utilized the majority of the time. However, in our patients it was not possible to use a single graft due to complex arch anatomy and short landing zones. Using a single graft in our cases could have led to stent migration and bird-beaking. With type 3 aortic arch anatomy, the endovascular graft repair is complicated by birdbeak phenomena [4]. In this situation, the proximal portion of the graft does not sit flush with the lesser curvature of the aortic arch. Due to high pressure blood flow from the ascending aorta, over time this will lead to invagination of the proximal portion, narrowing of the lumen, and migration of the graft. Type I endoleak, retrograde dissection, and graft collapse may occur due to a lack of apposition of the proximal stent-graft along the inner curvature of the aortic arch [10]. By utilizing multiple stent grafts with the turtlehead technique, a more rigid repair can conform to the aortic arch. This technique decreases the chances of encountering the negative sequelae of the bird-beaking due to precise deployment with good apposition to the lesser curvature of the aortic arch. Commercially available thoracic devices were initially derived from prototypes used years before in the infrarenal aorta [11, 12]. The “kilt technique” was developed in 2006 and describes a technique for unfavorable proximal landing zones of the infrarenal aorta. The authors describe their technique by first placing a cuff proximally within a reverse conical neck and then building downward with the exclusion stent graft device [13]. The turtlehead technique on the other hand, uses the final stent graft as a proximal cuff with 90% overlap for precision and elimination of bird-beaking along the natural curvature of the aortic arch. Due to the rigidity of this endograft construct, the chances of migration are reduced.

IMH is a type of aortic dissection. The vasa vasorum ruptures within the arterial wall causing weakness and impending rupture. Thoracic ballooning should be avoided in patients with severe aortic pathology where there is a substantial risk of causing aortic rupture or creating fenestrations [14]. Therefore, in the second case illustrated in this article, angioplasty was not an option to eliminate the bird-beak and using the turtlehead technique was the only endovascular option available. The mechanics of this technique are reliable because of the dominant amount of overlap within the confined area portion of the previous stent graft. There is no room for the final “turtlehead” graft to move or jump. Therefore, the most proximal portion of the stent graft is allowed to precisely flair open with wall apposition. The technique also eliminates the need to force positive pressure on the super stiff wire to attempt to achieve wall apposition on the outer curve. By placing forward tension on the deployment wire, the chances of unintended deployment movement increases. Although placing multiple grafts for difficult aortic pathology repair increases overall costs of the procedure, there are potential long-term benefits including less complications and compounding costs from re-interventions.

Follow-up imaging at four-month post procedure of our TAA patient demonstrated no migration with total exclusion of the aneurysm and preservation of the great vessels (Figure 24).

 

Conclusions

To date, we have utilized this stent stacking "turtlehead" method for multiple patients without complications related to stent migration, infolding or endoleak. Turtleheading provides an interesting technical strategy to overcome an unfavorable morphological situation at the proximal landing zone. The authors understand the financial burden of this technique and the cost of using multiple endografts. This technique offers a special niche for repair when the aortic arch anatomy is not conducive to other standard repair methods.


References

  1. Kuzmik GA, Sang AX , Elefteriades JA: Natural history of thoracic aortic aneurysms. J Vasc Surg. 2012, 56:565-571. Accessed: January 10, 2018: https://www.jvascsurg.org/article/S0741-5214(12)01039-7/fulltext. 10.1016/j.jvs.2012.04.053
  2. Madhwal S, Rajagopal V, Bhatt DL, Bajzer CT, Whitlow P, Kapadia SR: Predictors of difficult carotid stenting as determined by aortic arch angiography. J Invasive Cardiol. 2008, 20:200-204. Accessed: January 10, 2018: https://www.ncbi.nlm.nih.gov/pubmed/18460700.
  3. Bockler D, Muller-Eschner M, von Tengg-Kobligk H, Bischoff MS: Conformability in aortic type B dissection. Endovasc Today. 2014, 2:8-11. Accessed: January 10, 2018: https://evtoday.com/2014/02/supplement/conformability-in-aortic-type-b-dissection/.
  4. Findeiss LK, Cody ME: Endovascular repair of thoracic aortic aneurysms. Semin Intervent Radiol. 2011, 28:107-117. Accessed: January 10, 2018: https://www.thieme-connect.de/DOI/DOI?10.1055/s-0031-1273945. 10.1055/s-0031-1273945
  5. Chiesa R, Melissano G, Marrocco-Trischitta MM, Civilini E, Setacci F: Spinal cord ischemia after elective stentgraft repair of the thoracic aorta. J Vasc Surg. 2005, 42:11-17. Accessed: January 10, 2018: https://www.jvascsurg.org/article/S0741-5214(05)00563-X/fulltext. 10.1016/j.jvs.2005.04.016
  6. Feezor RJ, Martin TD, Hess PJ, et al.: Risk factors for perioperative stroke during thoracic endovascular aortic repairs (TEVAR). J Endovasc Ther. 2007, 14:568-573. Accessed: January 10, 2018: http://journals.sagepub.com/doi/10.1177/152660280701400420. 10.1177/152660280701400420
  7. Gillen JR, Schaheen BW, Yount KW, et al.: Cost analysis of endovascular versus open repair in the treatment of thoracic aortic aneurysms. J Vasc Surg. 2015, 61:596-603. Accessed: January 10, 2018: https://www.jvascsurg.org/article/S0741-5214(14)01773-X/pdf. 10.1016/j.jvs.2014.09.009
  8. Elefteriades JA: Natural history of thoracic aortic aneurysms: indications for surgery, and surgical versus nonsurgical risks. Ann Thorac Surg. 2002, 74:1877-1880. Accessed: January 10, 2018: https://www.annalsthoracicsurgery.org/article/S0003-4975(02)04147-4/fulltext. 10.1016/S0003-4975(02)04147-4
  9. Conrad MF, Ergul EA, Patel VI, Paruchuri V, Kwolek CJ, Cambria RP: Management of diseases of the descending thoracic aorta in the endovascular era: a Medicare population study. Ann Surg. 2010, 252:603-610. Accessed: January 15, 2018: 10.1097/SLA.0b013e3181f4eaef
  10. Böckler D, Schumacher H, Ganten M, et al.: Complications after endovascular repair of acute symptomatic and chronic expanding Stanford type B aortic dissections. J Thorac Cardiovasc Surg. 2006, 132:361-368. Accessed: January 15, 2018: https://www.jtcvs.org/article/S0022-5223(06)00726-4/fulltext. 10.1016/j.jtcvs.2006.02.056
  11. Volodos NL, Karpovich IP, Shekhanin VE, Troian VI, Iakovenko LF: A case of distant transfemoral endoprosthesis of the thoracic artery using a self-fixing synthetic prosthesis in traumatic aneurysm (Article in Russian). Grudn Khir. 1988, 6:84-86.
  12. Dake MD, Miller DC, Mitchell RS, Semba CP, Moore KA, Sakai T: The “first generation” of endovascular stent-grafts for patients with aneurysms of the descending thoracic aorta. J Thorac Cardiovasc Surg. 1998, 116:689-704. Accessed: January 15, 2018: https://www.jtcvs.org/article/S0022-5223(98)00455-3/fulltext. 10.1016/S0022-5223(98)00455-3
  13. Szaniewski K, Biernacka M, Walas RL Zembala M: Predeployed aortic extension cuff (kilt) in EVAR with hostile neck anatomy using Endurant II system: preliminary results. Polish J Thorac Cardiovasc Surg. 2016, 13:334-339. Accessed: January 15, 2018: https://www.termedia.pl/Journal/-40/Streszczenie-29004. 10.5114%2Fkitp.2016.64876
  14. Nation DA, Wang GJ: TEVAR: Endovascular repair of the thoracic aorta. Sem Intervent Radiol. 2015, 32:265-271. Accessed: January 15, 2018: https://www.thieme-connect.de/DOI/DOI?10.1055/s-0035-1558824. 10.1055%2Fs-0035-1558824
Technical report
peer-reviewed

Turtleheading the Tough Aortic Necks! A Novel Endovascular Method to Avoid Bird-beaking, Invagination, and Stent Migration in Thoracic Aortic Grafts


Author Information

Preston Hood Corresponding Author

Interventional Radiology, University of Florida College of Medicine, Jacksonville, USA

Maharshi Patel

Department of Interventional Radiology, University of Florida College of Medicine, Jacksonville, USA

Amanda Johnson

Department of Interventional Radiology, University of Florida College of Medicine, Jacksonville, USA

John Pirris

Cardiothoracic Surgery, University of Florida College of Medicine, Jacksonville, USA

Jerry Matteo

Radiology, University of Florida College of Medicine, Jacksonville, USA


Ethics Statement and Conflict of Interest Disclosures

Human subjects: Consent was obtained by all participants in this study. Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue. Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following: Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work. Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work. Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.


Technical report
peer-reviewed

Turtleheading the Tough Aortic Necks! A Novel Endovascular Method to Avoid Bird-beaking, Invagination, and Stent Migration in Thoracic Aortic Grafts


Figures etc.

SIQ
-
RATED BY 1 READER
CONTRIBUTE RATING

Scholary Impact Quotient™ (SIQ™) is our unique post-publication peer review rating process. Learn more here.