Calibration of Force/Torque and Acceleration for an Independent Safety Layer in Medical Robotic Systems



Abstract

Background: Most medical robotic systems require direct interaction with the robot. Force-Torque (FT) sensors can easily be mounted to the robot. However, an accurate FT control requires the current robot position to compute the spatial orientation of the sensor for gravity compensation.

Methods: We developed an independent safety system, named FTA sensor, which is based on an FT sensor and an accelerometer. With a calibration of accelerations to the FT coordinate frame, the current spatial orientation of the sensor is computed.

Results: We found that the calibration of accelerations into the FT coordinate frame can be performed with a median rotational error of 3.5°. The median error for gravity compensation based on accelerations was 0.3N and 0.04Nm for forces and torques, respectively.

Conclusion: By combining accelerations with force-torque readings, the FTA sensor works independently from robot input. Furthermore, the accuracy of the FTA sensor is sufficient for the purpose of medical robotic systems.

Introduction

Robotic assistance systems are more and more important for medical applications [1]. For recent developments on medical and surgical robotics see, e.g., [2-4]. Even though the applications and specifications differ, for many systems industrial robots are preferred instead of fully new designs [5-6]. The industrial robot is then adapted to the specific requirements of the applications. This adaptation is usually done in software, e.g. on the robot controller.

In neurosurgery, most robotic systems have been industrial robots [7-9]. Current developments are still based on industrial robot designs [10-12]. However, recent approaches also consider fully new designs, such as the MARS robot for stereotaxy [13].

For Transcranial Magnetic Stimulation (TMS), a magnetic coil is placed on the patient’s head for non-invasive brain stimulation. Recently, different robotic systems for TMS have been developed. A recent development introduces a specialized c-shaped robot design [14-16]. All other current robotic TMS-systems are based on industrial robots [17-21], including the commercially available TMS-robot SmartMove (Advanced Technology B.V., Enschede, The Netherlands).

For some medical robotic systems, force-torque sensors are mounted between robot end effector and tool [22-23]. These sensors are mainly used for haptic feedback, pressure control and/or user interaction. For accurate force and torque detection during operation, the tool’s weight-related forces and torques must be subtracted. As this impact changes, depending on the spatial orientation due to gravity, the spatial orientation of the sensor must be known. Commonly, this is done by using the current robot end effector position. Besides additional latencies, the communication with the robot controller is mostly done in software and the computation is not independent of the robot [22].

Therefore, we propose an independent safety system that is easily integrable in the existing systems and adds an additional safety layer to these systems. It is based on a force-torque (FT) sensor which is 

 

combined with an inertial measurement unit (IMU). An embedded system instantaneously triggers the robot emergency stop in case of an error or collision. As the key feature, the embedded system provides gravity compensation independently from robot input in real-time using the acceleration recordings.

In this paper, we present the idea of combining acceleration measurements with an FT-sensor for independence from robot input. We also address the issue of calibration of IMU to FT sensor and briefly describe the system’s setup and implementation. Besides evaluation of the calibration, we further show that the use of acceleration recording is sufficient for gravity compensation for medical robotic systems.

Materials & Methods

Common principle of gravity compensation

A force-torque (FT) sensor is capable of measuring forces and torques in the three spatial axes in real-time and high resolution. With such a sensor mounted to the robot end effector, we are able to detect impacts on the mounted tool. Due to gravity, the tool’s weight affects the sensor. To measure and detect impacts, e.g., user interaction or a collision, with the sensor, we must compensate for the tool weight. By changing spatial orientation of tool and sensor, the influence of the weight on the recordings changes.  Hence, we must consider the gravity compensation depending on the current robot orientation \(^{R}\xi_{E}\). Accordingly, we must know the transform ETFT from robot end effector to the sensor. This principle is illustrated in Figure 1.

When we mount the tool to the sensor and record the current force \(F=(F_x,F_y,F_z)'\), we can estimate the tool’s zero force \(F_0\) which is based on the force magnitude \(f=\left \| F \right \|_2\):

\(F_0=\begin{pmatrix} 0\\ 0\\ -f \end{pmatrix}\)

Hence, we can calculate the expected force \({F}'\) due to gravity and robot orientation:

\({F}'=(^{E}\xi_{FT})^{-1}\cdot (^{R}\xi_E)^{-1}\cdot F_0.\)

By subtracting \({F}'\) from the current force recording \(F\), we can calculate the applied force to the tool:

\(\widetilde{F}=F-{F}'.\)

In addition to forces, also torques due to gravity affect the recordings. For these torques \({M}'\) and the tool’s centroid \(s\) we have:

\({M}'={F}'\times s.\)

We subtract \({M}'\) from the current torque reading \(M\) to obtain the applied torque:

\(\widetilde{M}=M-{M}'.\)

To calculate the applied forces and torques the current robot orientation is required. In the presented

method above, the robot orientation \(^{R}\xi_{E}\) is fed in from the robot [22, 24].

Combining acceleration with force-torque

In contrast, an inertial measurement unit (IMU) can measure accelerations relative to gravity acceleration. Hence, the IMU is able to measure the gravity direction in relation to the IMU at rest. By combining such an IMU with an FT sensor, we can use the accelerations for gravity compensation. The combination of both sensors will be called FTA sensor. In contrast to FT sensors, IMUs are available as integrated circuits. As both, IMU and FT sensor, have their specific coordinate frame, we must perform a calibration between both the sensors. Thereby, we get the transformation matrix \(^{FT}\xi_{IMU}\) to convert the accelerations A from the IMU to the FT coordinate system:

\(A_{FT}=^{FT}\xi_{IMU}\cdot A_{IMU}.\)

Now, we can use the accelerations to compensate for gravity. We calculate the expected force F0 for the current orientation with:

\({F}'=A_{FT}\cdot f.\)

We estimate the applied forces \(\widetilde{F}\) and torques \(\widetilde{M}\) corresponding to the above presented equations but with usage of Eq. (7) instead of Eq. (2). This way, robot input is not required for computing the spatial orientation of the sensor. Hence, it operates independently.

We use an embedded system (ES) for implementation of the calculations in real-time [25]. Figure 2 shows the communication setup for the embedded system. The ES directly reads the input from the IMU and the FT sensor and performs the calculations and triggers the robot’s emergency stop in case of an error.

Setup and circuit board

We use a K6D force-torque sensor (ME Systeme, Heringsdorf, Germany) and integrate the sensor into a specific casing. This casing allows for easy mounting to the robot end effector. Besides the FT sensor, the casing contains and protects the circuit board, including the IMU.

We mount the FTA sensor to an Adept Viper s850 industrial robot (Adept Technology, Inc., Livermore, CA, USA) as shown in Figure 3. We pass the communication and power supply cable through the robot’s internal user communication interface. This way, intertwining of the cable with the tool or articulated arm is avoided.

The cicuit board consists of the IMU (LIS3LV02DQ; STMicroelectronics N.V., Amsterdam, The Netherlands) as the 3D accelerometer and a relay for connection to the emergency stop. The IMU is a three axes linear accelerometer with a measurement range of up to \(\pm 6G\), with \(1G=9.81\frac{m}{s^{2}}\)

Furthermore, an analog-digital converter (ADC) is located on the board for reading the voltages from the FT sensor. Also, the board consists of a direct current converter for power supply. As microprocessor, we use an Atmel AT32 with a bandwidth of 32 bits and a processor clock rate of 60MHz.

Calibration of IMU to FT sensor

As IMU and FT sensor are located in the same casing, a coarse knowledge of their coordinate systems exists. However, for our application, an accurate transformation is required. Thus, a calibration of IMU to FT sensor is mandatory.

Once the FTA sensor is installed to the robot, we use a full circular motion in joint 4 of the articulated arm to perform calibration. For the circular motion, the angle values are used with the measured acceleration and Joint 5 is set to \(45^{\circ}\) to allow for non-zero measurements in all spacial axes. For

calibration, we mount a weight to the FT sensor.

For each spatial axis and for each modality (force, torque, acceleration), we calculate a cosine fit using:

\(a_{1}\cos (\gamma +b_{1})+c_{1};\)         \(\gamma \in \left [ -\pi, \pi \right ],\)

with \(l=F_x,F_y,F_z,M_x,M_y,M_z,A_x,|A_y,A_z.\) In this case, the parameter \(c_l\) describes the offset for forces, torques and accelerations. By comparison of the phase angle \(c_l\) between forces \(F\) and accelerations \(A\), we can compute the transform \(^{FT}\xi_{IMU}\) between FT sensor and IMU. As the translational shift of the IMU is meaningless, the transform only consists of a rotational matrix.

Due to the system setup (cf. Figure 4), we have a coarse knowledge of the orientation of IMU and FT sensor:

\(\overrightarrow{e}_{FT_{x}}\approx -\overrightarrow{e}_{A_{z}}\)

\(\overrightarrow{e}_{FT_{y}}\approx \overrightarrow{e}_{A_{y}}\)

\(\overrightarrow{e}_{FT_{z}}\approx \overrightarrow{e}_{A_{x}},\)

where \(\overrightarrow{e}\) denotes the corresponding unit vector. Figure 5 illustrates this relationship with recorded force and acceleration measurements. Also, the cosine fit for each modality is shown. Consequently, we know that a rotation of \(\approx-90^{\circ}\) around the y-axis is needed to transform accelerations into the

FT-sensor coordinate frame. Also, the other phase angles must be adapted, resulting in the following equation:

\(^{FT}\xi_{IMU}\approx R_z(0)\cdot R_x(0)\cdot R_y(\frac{-\pi}{2}),\)

where \(R_{y}\) describes a rotation around the y-axis, \(R_{z}\) and \(R_{x}\) around z- and x-axis, respectively.

Using the phase angles \(b_{l}\), the equation can now be refined as:

\(^{FT}\xi_{IMU}=R_z(b_{F_{z}}-b_{A_{x}})\cdot R_x(b_{F_{x}}-b_{A_{z}})\cdot R_y(\frac{-\pi}{2}-(b_{F_{y}}-b_{A_{y}})).\)

Note that the Equations (12) and (13) can be easily adapted to any other system setup. The rotational matrices must be changed in accordance with the specific setup. Also, we are using the calibration matrix C which converts the voltage readings from the FT sensor into forces and torques. As a result, we use

\(F_{user}=\begin{pmatrix} (C\cdot V)_1-c_{F_{x}}-(^{FT}\xi_{IMU}\cdot A)_x\cdot F_G\\ (C\cdot V)_2-c_{F_{y}}-(^{FT}\xi_{IMU}\cdot A)_y\cdot F_G\\ (C\cdot V)_3-c_{F_{z}}-(^{FT}\xi_{IMU}\cdot A)_z\cdot F_G \end{pmatrix}, and\)

\(M_{user}=\begin{pmatrix} (C\cdot V)_4-c_{M_{x}}-((^{FT}\xi_{IMU}\cdot A\cdot F_G)\times s)_x\\ (C\cdot V)_5-c_{M_{y}}-((^{FT}\xi_{IMU}\cdot A\cdot F_G)\times s)_y\\\ (C\cdot V)_6-c_{M_{z}}-((^{FT}\xi_{IMU}\cdot A\cdot F_G)\times s)_z \end{pmatrix}\)

to estimate the gravity compensated forces \(F_{user}\) and torques \(M_{user}\), based on the voltage readings \(V\), the accelerations \(A\) and the tool’s gravity force \(F_G\) and centroid \(s\). Note that different methods for estimating the tool’s weight and centroid exist, e.g. [24] for a TMS coil.

Evaluation

Calibration

First, we evaluate the accuracy of the calibration from IMU to FT sensor. Therefore, we perform the presented calibration method with two different FT sensors and two IMUs (including circuit board with ES), resulting in a total of four FTA sensors. For each FTA sensor, we perform three sets of calibrations with 20 calibrations in a 15-min-interval. We therefore have 60 calibrations of IMU to FT sensor for each FTA sensor that we use for evaluation.

Quality of the fit

As the calibration is based on fitted values (cf. Equation (8)), the quality of the fit is essential for the accurateness of the calibration. Therefore, we estimate for each recording of each modality the absolute distance to the fitted curve.

Calibration error

For calculating errors of the calibration, we first transfer the recorded accelerations \(A_{IMU}\) into the FT coordinate frame by applying the computed transformation matrix \(^{FT}\xi_{IMU}\) (cf. Equation 6). We fit the transferred accelerations to a cosine with the formula from Equation (8). We compare the phase angles of the forces (estimated during calibration) to the phase angle of the transferred accelerations \((A_{FT})\) and compute the error for each spatial axis by applying the inverse sine to the phase difference.

Stability of calibration

For calculating the stability of the calibration, two calibration results \(T_{1}\) and \(T_{2}\) are used. To compare the difference between these two, we use

\(T_{e_{1}}=T_1\cdot T_2^{-1}\) and \(T_{e_{2}}=T_2\cdot T_1^{-1}.\)

where \(T_{e_i}\) are rotational matrices. The stability is now expressed as the computed rotational error \(E_{rot}\) as

\(E_{rot}=\frac{1}{2}(\left | \theta_1 \right | + \left | \theta_2 \right |),\)

using the axis-angle (i.e., \((a_i, \theta_i)\)) representation of the matrices \(T_{e_i}\).

The use of both relationships \(T_{e_1}=T_1\cdot T_2^{-1}\) and \(T_{e_2}=T_2\cdot T_1^{-1}\) is necessary since the matrices \(T_1\) and \(T_2\) may be non-orthogonal. Consequently, since we do not wish to privilege one frame of reference, the average of the errors is used. This, and the way of computing the rotational error, is in line with standard approaches for hand-eye calibration [26-27]. Note that, as the calibration of IMU to FT only consists of a rotational part, no translational error is estimated.

Gravity compensation

To estimate the goodness of the independent gravity compensation based on accelerations, we mounted a weight to the sensor and estimated the tool’s weight and centroid [24]. We used these parameters for gravity compensation (Equations (14) and (15)). We now moved the robot randomly within all spatial axes and recorded the gravity compensated forces and torques from the FTA sensor. In this way, we collected roughly 20;000 data points which we used for evaluation.

Results

Calibration

Quality of the fit

Figure 6 shows the overall cosine fitting quality used for calibration as boxplots. The median deviations for forces were 0:14N, 0:11N and 0:15N for the three spatial axes. For torques, it was 0:0034Nm, 0:0023Nm and 0:0017Nm, respectively. The median deviations for the accelerations were 0:016G, 0:027G and 0:022G, respectively. Due to noise, we were not able to perform a valid cosine fitting in two recordings. Therefore, these two recordings were excluded from further analysis.

Calibration error

The median calibration error, was \(3.4^{\circ}\) for the x-axis and \(3.5^{\circ}\) and \(1.6^{\circ}\) for the y- and z-axis, respectively. Figure 7 shows these results as boxplots.

Stability of the calibration

In total, we evaluated the stability of the calibration on almost 7,000 combinations of calibration results. Figure 8 shows the results as a boxplot. The median deviation was \(0.89^{\circ}\). For the sensors 1 and 3, the median error was even below \(0.7^{\circ}\).

Gravity compensation

Figure 9 shows the error of the gravity compensated forces and torques. On average, the error for forces was in the range of 0.3-0.4N for each spatial axis. For torques, the average error was in the range of 0.02-0.045Nm. Note that the used weight corresponded to approximately 0.7Kg.

Discussion

We presented the use of acceleration measurements in combination with an FT sensor to perform gravity compensation independent from the robot. The necessary calculations for combining both sensors can be performed with an embedded system in real-time [25]. In this way, it acts as an independent safety-layer for medical robotics systems.

We have shown that the required calibration of the accelerations to the force/torque sensor coordinate frame can be done with a median error of roughly \(3.5^{\circ}\). However, there have been some recordings with a larger fitting error due to noise in the measurements. As the calibration is only

required once for each FTA sensor, we are able to repeat and extensively validate the calibration result. For instance, we can use the fitting error to validate if the quality of the measurements is poor. In case of noise, we will repeat the recordings to minimize the error. Therefore, it will be possible to perform a final calibration of the FTA sensor with a calibration error below \(2^{\circ}\).

Our evaluation suggests that the presented calibration method produces stable results. The median deviation between two calibration matrices was \(0.89^{\circ}\).

Besides these evaluations on the calibration itself, our practical test shows that the gravity compensation based on accelerations is sufficient for the application. The median error was roughly 0.3N for the force readings and approximately 0.03-0.04Nm for the torque readings. The maximum errors were below 1.25N and 0.13Nm for forces and torques, respectively.

For the robotized TMS system, the used contact pressure is in the range of 2-5N [22]. For user interaction with the robot using hand-assisted positioning, only forces larger than 2N and torques larger than 0.5Nm are taken into account to move the robot. Therefore, the presented gravity compensation is sufficient and applicable for the purpose of robotized TMS.

By combining accelerations and force-torque measurements with the data processing of a real-time embedded system, we can use the presented method for real-time monitoring of the robot. In case of an error in the computation cycle, from reading of the measurements to the computation of the gravity compensated forces and torques, the FTA sensor can instantaneously trigger the robot’s emergency stop [25]. The FTA sensor itself is easily mountable between robot end effector and tool. For safety monitoring, the used software does not have to be changed or adapted. The sensor runs independently from robot and software. Note that the presented gravity compensation based on accelerations also works in case the robot is positioned skewly (not aligned with the gravity). When

using gravity compensation based on robot input, the rotation of the robot with respect to the direction of gravity must be taken into account.

Conclusions

In conclusion, the FTA sensor measures applied forces and torques independent of the robot by combining an FT sensor with an accelerometer. The required computations can be done on an embedded system which can be added to the circuit board next to the accelerometer. These computations can be performed in real-time enabling the embedded system running a real-time monitoring cycle to control the robot. Therefore, it can stop the robot instantaneously in case of a collision or error to prevent patient and/or operator from serious harm. For the robotized TMS system, it is an important safety feature and a prerequisite for its safe clinical application. Our practical tests have shown, that the gravity compensation based on accelerations is sufficient and applicable for medical robotics systems.


References

  1. Taylor RH: A Perspective on Medical Robotics. Proceedings of the IEEE. 2006, 94:1652 –1664.
  2. Bozovik V: Medical Robotics. InTech, Vienna, Austria; 2008.
  3. Baik SH: Robot Surgery. InTech, Vukovar, Croatia; 2010. 10.5772/6897
  4. Lanfranco AR, Castellanos AE, Desai JP, Meyers WC: Robotic surgery: a current perspective. Annals of Surgery. 2004, 239:14-21.
  5. Adler JR Jr, Chang SD, Murphy MJ, Doty J, Geis P, Hancock SL: The CyberKnife: A frameless robotic system for radiosurgery. Stereotactic and Functional Neurosurgery. 1997, 69:124–128.
  6. Schweikard A, Bodduluri M, Adler JR Jr: Planning for camera-guided robotic radiosurgery. IEEE Transactions on Robotics and Automation. 1998, 14:951–962.
  7. Kwoh YS, Hou J, Jonckheere E, Hayati S: A robot with improved absolute positioning accuracy for CT guided stereotactic brain surgery. IEEE Trans Biomed Eng. 1988, 35:153–160.
  8. Benabid A, Cinquin P, Lavalle S, Le Bas J, Demongeot J, De Rougemont J: Computer-driven robot for stereotactic surgery connected to CT scan and magnetic resonance imaging. Stereotactic and Functional Neurosurgery. 1987, 50:153–154.
  9. Drake JM, Joy M, Goldenberg A, Kreindler D: Computer and robot-assisted resection of thalamic astrocytomas in children. Neurosurgery. 1991, 29:27-33.
  10. Medtech. Rosa neurosurgery robot. Montepllier, France: Medtech SAS. (2011). http://www.medtechsurgical.com/products/ROSA.
  11. Li QH, Zamorano L, Pandya A, Perez R, Gong J, Diaz F: The application accuracy of the Neu- roMate robot - A quantitative comparison with frameless and frame-based surgical localization systems. Computer Aided Surgery. 2002, 7:90–98.
  12. Morgan PS, Carter T, Davis S, et al.: The application accuracy of the Pathfinder neurosurgical robot. International Congress Series CARS 2003 Computer Assisted Radiology and Surgery Proceedings of the 17th International Congress and Exhibition. 2003, 561–567.
  13. Heinig M, Govela MF, Gasca F, et al.: MARS - Motor Assisted Robotic Stereotaxy System. Proceedings of the 5th International IEEE EMBS Conference on Neural Engineering. 2011. 334 – 337.
  14. Zorn L, Renaud P, Bayle B, Goffin L, Lebosse C, de Mathelin M, et al: Design and evaluation of a robotic system for transcranial magnetic stimulation. Biomedical Engineering, IEEE. 2012, 59:805-15.
  15. Lebosse´ C, Renaud P, Bayle B, de Mathelin M, Piccin O, Foucher J: A robotic system for automated image-guided transcranial magnetic stimulation. Life Science Systems and Applications Workshop, 2007. LISA 2007. IEEE/NIH. 2007. 55-58.
  16. Renaud P, Piccin O, Lebosse´ C, Laroche E, de Mathelin M, Bayle B, et al.: Robotic image-guided transcranial magnetic stimulation. Computer Assisted Radiology and Surgery (CARS), 20th International Congress. Osaka, Japan; 2006.
  17. Matthäus L: A robotic assistance system for transcranial magnetic stimulation and its application to motor cortex mapping. Universität zu Lübeck. 2008, Ph.D. thesis:
  18. Matthäus L, Trillenberg P, Bodensteiner C, Giese A, Schweikard A: Robotized TMS for motion compensated navigated brain stimulation. Computer Assisted Radiology and Surgery (CARS), 20th International Congress. Osaka, Japan; 2006.
  19. Matthäus L, Giese A, Wertheimer D, Schweikard A: Planning and analyzing robotized TMS using virtual reality. Studies in Health Technology and Informatics. 2006, 119:373–378.
  20. Lancaster JL, Narayana S, Wenzel D, Luckemeyer J, Roby J, Fox P: Evaluation of an image- guided, robotically positioned transcranial magnetic stimulation system. Human Brain Mapping. 2004, 22:329–340.
  21. Yi X, Bicker R. : Design of a robotic transcranial magnetic stimulation system. IEEE Conference on Robotics, Automation and Mechatronics. Singapore; 2010.
  22. Richter L, Bruder R, Schweikard A: Hand-assisted positioning and contact pressure control for motion compensated robotized transcranial magnetic stimulation. Int J Comput Assist Radiol Surg.. 2012, 7:845-52. 10.1007/s11548-012-0677-6
  23. Dombre E, Duchemin G, Poignet P, Pierrot F. Dermarob: A safe robot for reconstructive surgery. Robotics and Automation, IEEE Transactions on. 2003, 19:876-884.
  24. Richter L, Bruder R, Schlaefer A: Proper Force-Torque Sensor System for robotized TMS: Automatic Coil Calibration. Int J Comput Assist Radiol Surg; Proceedings of the 24th International Conference and Exhibition on Computer Assisted Radiology and Surgery (CARS’10). 2010, 5:S422–S423.
  25. Richter L, Bruder R: Design, implementation and evaluation of an independent real-time safety layer for medical robotic systems using a force-torque-acceleration (FTA) sensor. Int J Comput Assist Radiol Surg. 2013, 8:429-36. 10.1007/s11548-012-0791-5
  26. Strobl KH, Hirzinger G: Optimal Hand-Eye Calibration. Proceedings of the IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS 2006), Beijing, China. 2006. 4647-4653.
  27. Richter L, Ernst F, Schlaefer A, Schweikard A: Robust Robot-Camera Calibration for Robotized Transcranial Magnetic Stimulation. International Journal of Medical Robotics and Computer Assisted Surgery. 2011, 7:414–422.
Original article
peer-reviewed

Calibration of Force/Torque and Acceleration for an Independent Safety Layer in Medical Robotic Systems


Author Information

Lars Richter Corresponding Author

Institute for Robotics and Cognitive Systems, University of Luebeck; Graduate School for Computing in Medicine and Life Sciences, University of Lubeck, 23538 Lubeck, Germany

Ralf Bruder

Institute for Robotics and Cognitive Systems, University of Lubeck,, Institute for Robotics and Cognitive Systems, University of Lubeck,

Achim Schweikard

Institute for Robotics and Cognitive Systems, University of Luebeck, Institute for Robotics and Cognitive Systems, University of Lubeck


Ethics Statement and Conflict of Interest Disclosures

Human subjects: All authors have confirmed that this study did not involve human participants or tissue. Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue. Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following: Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work. Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work. Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.